phenomenon
principle
deal with
1. No power supply
1. The plug is not inserted or short
1. Insert the plug
2. Open fuse
2. Replace the fuse tube of the same model
3. The power switch is broken
3. Replace the power switch
2. The incubator does not open the cover
1. Low setting temperature
1. Adjust the set temperature
2. The probe is broken
2. Change the probe
3. Air leakage in operation room and sampling room
1. Poor sealing
1. Seal with soap water and seal with 704
2. Aging rubber ring
2. Replace the rubber band
4. No gas connection
1. The valve is broken
1. Replace the air valve
2. Leakage or fall off of the leather tube
2. Replace the special hose
3. Running out of gas cylinders
3. Replace the gas
5. The light is off
1. The starter or lamp does not light
1. Replace the starter or lamp
2. Rectifier burned out
2. Replace the rectifier
Shanghai Haozhuang Instrument Co., Ltd. specializes in the production of laboratory equipment, anaerobic incubators, drying ovens, various incubators, carbon dioxide cell incubators, ultrasonic cleaners, shakers, purification workbenches and other products. Welcome new and old customers to inquire!